FYCS-Z Wire et cable Bunched Igneus Test Equipment (Missa O Controller)
depictio producti
Aptum est aestimare facultatem institutionis verticalis fili fasciculi et funis vel optici ad supprimere flammam verticalem sub certis conditionibus divulgatam.
Standard
Parere cum GB18380.31-2022 "Combustionis probatio funerum sub flammarum conditionibus Pars III: Verticalis institutionem filum fasciculum et funem flammae verticalis testium machinam divulgatam", aequivalens cum IEC60332-3-10:2000.
Eodem tempore ut requisita Tabula IV GB/T19666-2019 "Principia Generalia Flammae Retardant et Refractionis Wire et Cable" vexillum.
GB/T18380.32--2022/IEC60332--3--21: 2015 "Combustio funerum electricorum et funerum opticorum sub flammae conditionibus Pars 32: Vertice inauguratus est filum fasciatum et funis flammae verticalis diffusae test categoriae AF/R".
GB/T18380.33--2022/IEC60332--3--22: 2015 "Combustio funerum electricorum et funerum opticorum sub flammarum conditionibus Pars 33: Vertice inauguratus est filum aggregatum et funem flammae verticalis divulgationis test categoriae A".
GB/T18380.35--2022/IEC60332--3--24:2015 "Combustio funerum electricorum et funerum opticorum sub flammarum conditionibus Pars 35: Vertice inauguratus est filum aggregatum et funem flammae verticalis expansum test categoriae C"
GB/T18380.36--2022/IEC60332--3--25: 2015 "Combustio funerum electricorum et funerum opticorum sub flammarum conditionibus Pars 36: Verticalis inauguratus filum fasciculum et funis flammae verticalis diffusae test categoriae D".
Apparatus Compositio
Combustio test thalamum, ratio electrica imperium, fons aer, ignitio fons massae fluunt systematis imperium (gas propansum et gas compressum aeris), scala ferri, ignis fabrica exstinguentis, emissio machinae purificationis, etc.
Technical Parameter
1.Working voltage: AC 220V±10% 50Hz,Power consumption: 2KW
2.The inlet and outlet air flow rate: 5000±200 L/min(adjustable)
3.Air fluxus et propanus fluunt mole fluunt systematis temperantiae reguntur.
4.Air principium: propane (0.1Mpa), aer (0.1Mpa), mos-possessores aeris principium.
5.Timing range: 0 ~ 60min (potest)
6.Anemometer measurement range: 0 ~ 30m/s, measurement accuracy: ±0.2m/s
7.Test dimensio cubiculi(mm): 2184(L) x 1156(W) x 5213(H)
Repletus ignium mineralibus resistens materia velit petra lanae, summa cum 1500 mm alta custodia tuta.
8.2 Combustio capitis cum Venturi mixer
9. Aer diverticulum ventilabrum est humili strepitus vorticis ventilabrum. PLC celeritatem ventilabrum per frequentiam convertentis moderatur, et vortex fluens metitur aerem volumen ad consequendum sessorium aeris subtilis voluminis imperium.
10. Inducta captura fan adoptat 4-72 anti-corrosionem ventilabrum cum aere volumine 5000m2/h.
11.Flue gas post-curatio aqua imbre pulvis remotionis turris instructa est cum volumine aeris processus 5000 m2/h
12. Uterque NITROGENIUM ignis exstinguentis et aquae aspergine ignis exstinguendi methodi instructi clientibus eligendi sunt.
13.Ad probationem:
Vexillum ferri rectilineum scalae dimensio(mm): 500(W) x 3500(H)
Scala ferri lata dimensio verticalis (mm): 800(W) x 3500(H)
14.Combustion dimensionis superficiei(mm): 257(L) x 4.5(W)
15.Touch screen imperium, intuitive et clarum, ignitionem electronicam, sincere automatarium.
16. USTOR PLC refrenat et operetur per tactum velum.
Test Fabrica
Test box: The experimental device should be a self-standing box with a width of 1000mm, a depth of 2000mm and a height of 4000mm. The bottom of the box should be 300mm above the ground. The perimeter of the test chamber should be sealed, the air from the bottom of the chamber from the front wall (150±10) mm to open a (800±20) mm x (400±10) mm air inlet into the box. A (300±30) mm x (1000±100) mm outlet should be opened at the back of the top of the chamber. Test chamber should be used on both sides of the heat transfer coefficient of about 0.7W.m-2.K-1 thermal insulation, the distance between the steel ladder and the back wall of the test chamber is (150±10) mm, and the bottom rung of the steel ladder is (400±5) mm from the ground. The lowest point of the cable sample is about 100mm from the ground.
-
Standard Venturi Blowtorch
-
Blowtorch Hole
-
Burner
-
Venturi mixer
1.Anemometer: celeritatem venti extra summum cubiculi testi metitur, si celeritas venti 8m/s excedit tunc experimentum agi non potest.
2. Temperatura probe: duo K-types thermocouples utrinque instructi sunt cistae test, si caliditas muri interioris minor est quam 5℃ vel superior quam 40℃, experimentum ferri non potest.
3.Air source: adopt touch screen controller, frequency conversion control inlet axial flow fan, can intuitively read out and control the gas flow through the air box for (5000±200) L/min, stable air flow rate during the test.
4. Post peractam probationem: Si specimen adhuc ardet, post unam horam incendium claudendi, aqua aspergine fabrica vel nitrogenis ignis artificium exstinguendi adhiberi potest ad ignem violenter sistendum, et peculiaris infundibuliformis est ad repurgandum. perditio.
5.Steel ladder type: width (500±5)mm standard steel ladder, width (800±10)mm wide steel ladder, material for SUS304 stainless steel tube.

Unus quisque pro vexillo et lato ferro scalis
Emissio Purificationis Fabrica
Collectio fumi et machinae fuliginis lotio: materia PP cum diametro 1500mm et altitudo 3500mm. Fumus turris collectio in tres partes dividitur: imbre fabrica, fumum et pulverem spargendi fabrica, ac fumum fabricam exhauriunt. Imbre fabrica: ad imbrem aquarum specialium materiarum colum praebendarum, ut speciales materias sparguntur ad fumum et pulverem eliquandum efficaciter. Fumus et pulvis colum notae: Filtratum bibendo materiam spargendi aquam, quae efficaciter fumum et pulverem spargere potest ut fumus emissus sit fumus albus. Clientes clientes instrumenti tutelae environmental iuxta condicionem addunt.
-
Fumus Collection Turris Schematic
-
Fumus turris collectio
-
Adductus in capturam fan
Ignitio Source
1.Ignition source type: including one or two band-type propane gas blowtorches and their matching flowmeters and Venturi mixers. The ignition surface is drilled with 242 flat metal plates with a diameter of 1.32mm. The center distance of these holes is 3.2mm, arranged in three rows in a staggered arrangement, each row is 81, 80 and 81, distributed in the nominal size is 257×4.5mm. In addition, a row of small holes are opened on both sides of the flame board, and this guide hole can maintain the stable combustion of the flame.
2.Ignition source location: The torch should be placed horizontally, (75±5) mm from the front surface of the cable sample, (600±5) mm from the bottom of the test chamber, and symmetrical to the axis of the steel ladder. The flame supply point of the blowtorch should be located in the center between the two crossbeams of the steel ladder, and at least 500mm away from the lower end of the sample. The center line of the blowtorch system should be approximately the same as the center line of the steel ladder.
-
Vortex Flow Meters for
Diligenter Imperium Inlet Aeris Volume -
Vortex aeris diverticulum fan
Missa O Controller
Massa moderatoris fluere adhibetur pro certa mensura et potestate massae fluxus gasi. Missae metra fluunt altae accurate lineamenta, iterabilitas bona, responsio ieiunium, initium mollis, stabilitas et fides, ampliatio pressionis operativae latae. Cum connexionibus internationalibus vexillum, facile est operari et uti, in quolibet loco institui potest, et facile coniungi cum una PC pro automati potestate.
Missa fluxus moderatoris technici parametri:
1.Accuracy: ±2% F.S
2.Linearity:±1% F.S
3.Repeat accuracy:±0.2% F.S
4.Response time:1 ~ 4 sec
5.Pressure resistentia: III Mpa
6.Working environment:5 ~ 45℃
7.Input exemplar: 0-+5v